Why my metabolite concentrations deviate from predicted concentrations?
Metabolite intensities (peak areas) depend on several factors, such as ion suppression effects, caused by the sample’s matrix, metabolite dilution or extraction effects, e.g. the metabolite extraction efficiency from a cell pellet, or the ratio of ...
What are the common sources of fatty acids contamination in the LC system?
Fatty acids (FA) are substances, which are common contaminations in LC-MS systems and they “like to remain” in the system. FA are part of the most biological samples and can be easily broguht into the system. e.g., solvent bottles. Even after a ...
What happens if I use the kit after it expires?
All kit components requiring a refrigeration are subject to expiration (light blue box). These are the Vial Box and the Kit Plate. After passing the expiration date, the kit can still be used. However, the proper performance cannot be guaranteed. ...
Missing concentration values in Results
Whenever a concentration cannot be calculated, a specific status is shown in WebIDQ > Results. Due to analytical reasons, concentration values may not be available, and will be represented by these statuses: NA Empty values are replaced with NA NaN ...
What kind of plasma anticoagulants are compatible with biocartes kits?
biocrates kits were validated with EDTA and heparin plasma only. Citrate plasma may also be used, but performance may be reduced. The most important aspect is to ensure that all plasma samples within one study are collected in the same way and using ...