Change OP of kit plate

Change OP of kit plate

If another OP was linked with a kit plate in WebIDQ than required for the used mass spectrometer system, the OP may be changed after running a kit. An example is given for the platform SCIEX 5500+ and the Quant 500 kit.
– The OP for SCIEX 6500+ was linked with the kit plate, MXP500L-0-5713.
– The OP for SCIEX 5500+ is MXP500L-0-5813.

It is required to change the OP in WebIDQ and to rename all MS data files.

Change OP
1. In LIMS > Projects select the corresponding plate run, e.g. 1052365870-1, and click Delete measurements.

2. Change the OP from MXP500L-0-5713 to MXP500L-0-5813.



Rename all MS data files
1. Make copies of the corresponding MS data files, e.g. all WIFF and SCAN or RAW files of the plate run.

2. Rename the copied MS data files. For example,
Original file name | for MXP500L-0-5713:
MXP500L-0-5713_1052365870_01_0_1_1_01_10000001.wiff
New file name | for MXP500L-0-5813:
MXP500L-0-5813_1052365870_01_0_1_1_01_10000001.wiff
InfoFor file renaming a bulk rename tool may be used.

Data quantification
It is required to integrate, calibrate, and validate the plate run again.
1. In Quantification > Integration import the renamed MS data files.
2. Follow the procedures described in the WebIDQ user manual in section Quantification.
InfoAfter changing an OP, it is required to check all retention time (RT) windows. To import RT windows which were previously defined, a RT window export file (CSV export from WebIDQ) may be used.See section Export integration windows.

    • Related Articles

    • Change plate layout after kit run

      Change plate run after kit run. Please follow the instructions in the pdf attachment.
    • How to change the sample volume in WebIDQ?

      The sample volume is defined in WebIDQ. Itis the volume which was pipetted to each filter spot of the kit plate. The standard value is 10 µL. This procedure describes, how to change thesample volume in WebIDQ. To update sample volume added to the ...
    • Recreate plate runs and restore sample information

      Import plate run(s) This feature was previously named Import worklist – Plateruns can be recreated using the corresponding MS datafiles. – Please find a description in theuser manual, Import platerun(s). – Data from the LC and FIA parts maybe ...
    • Cannot access the project or plate run?

      If you are not able to access a project or a plate run in WebIDQ, it is because of the project team restriction. Ask the project owner or WebIDQ admin to grant you the rights by adding in the project team list.
    • How to delete, insert or swap samples in WebIDQ?

      To copy, cut, delete, or select all, right click on samples. To select multiple samples, press either Ctrl or shift while selecting the samples. To insert additional samples or paste a copied or cut sample, right click on empty wells. Swap ...