How is LOD calculated?

How is LOD calculated?

The limit of detection (LOD) for each metabolite within a kit run is calculated from the concentrations in the zero samples based on either the standard deviation (StdDev-based) or median (Median-based).

  1. StdDev-based: LOD is calculated as median concentration of the zero samples concentrations plus three times the standard deviation of the zero samples concentrations.
LOD = median background signal + 3 × standard deviation
  1. Median-based: LOD is calculated as three times the median concentration of the zero samples.
LOD = median background signal × 3

For detailed description, please refer to the WebIDQ user manaul.

    • Related Articles

    • Migration wizard and several other settings are not accessible in WebIDQ account.

      The migration wizard is accessible only with the administrative privileges. If the user don't have admin rights in WebIDQ then Migration wizard in WebIDQ account will not be visible. There are several other functionalities exclusively available for ...
    • How to delete, insert or swap samples in WebIDQ?

      To copy, cut, delete, or select all, right click on samples. To select multiple samples, press either Ctrl or shift while selecting the samples. To insert additional samples or paste a copied or cut sample, right click on empty wells. Swap ...
    • Recreate plate runs and restore sample information

      Import plate run(s) This feature was previously named Import worklist – Plate runs can be recreated using the corresponding MS datafiles. – Please find a description in theuser manual, Import plate run(s). – Data from the LC and FIA parts maybe ...
    • Missing concentration values in Results

      Whenever a concentration cannot be calculated, a specific status is shown in WebIDQ > Results. Due to analytical reasons, concentration values may not be available, and will be represented by these statuses: NA Empty values are replaced with NA NaN ...
    • How to change the sample volume in WebIDQ?

      The sample volume is defined within WebIDQ, representing the precise amount of liquid that is pipetted onto each filter spot of the kit plate. By default, the standard sample volume is set to 10 µL. However, in certain cases, it may be necessary to ...