How to overcome “no intercept”
Whenever the status “No Intercept” or “infinity
concentration” is displayed instead of an analyte concentration change in the
quantification process can be performed.
Measured area ratio values which do not intercept with a quadratic
calibration curve, result in “No Intercept” or “infinity concentration”.
Generally, this can be considered as >ULOQ. To get value for these analytes,
the curve fit type may be switched from quadratic to linear. If several plates
are analyzed together, it is recommended to change the calibration curve of the
affected analytes to linear for all plates.
Customized quantification
Due to large concentration ranges in samples, metabolite
concentrations may be above the upper limit of quantification
(ULOQ). If a quadratic regression is used for quantification, values above the
ULOQ may cause the quantification status “no intercept”. Consequently, an adaptation
of the quantitation process may be necessary.
- Perform
the quantification according to the kit user manual. In WebIDQ go to Quantification > Validation or Results and check for which analytes
the status “no intercept” is displayed.
- Check
for which analytes the status “no intercept” is given. Check if the status “no intercept”
is given for only some or the majority of samples. If the majority of samples
is affected, set the regression that is used for quantification to linear and
exclude upper calibration standard level(s), which do not match the regression
line. Do so for all affected analytes. Please find an example below.

Related Articles
Missing concentration values in Results
Whenever a concentration cannot be calculated, a specific status is shown in WebIDQ > Results. Due to analytical reasons, concentration values may not be available, and will be represented by these statuses: NA Empty values are replaced with NA NaN ...
Why are QC accuracies consistently low or high for all analytes?
If the QC accuracies are consistent for all measured analytes, but differ significantly from the expected values (i.e. all metabolites for any particular QC sample are at 50% accuracy), the following may have occurred: 1. Pipetting issue: More or ...
Change OP of kit plate
If another OP was linked with a kit plate in WebIDQ than required for the used mass spectrometer system, the OP may be changed after running a kit. An example is given for the platform SCIEX 5500+ and the Quant 500 kit. – The OP for SCIEX 6500+ was ...
Change plate layout after kit run
Change plate run after kit run. Please follow the instructions in the pdf attachment.