Maximum unknown sample volumes for MxP Quant 1000 kit?
If you wish to add more sample volume on a filter of the test plate, please note that extra drying steps are required to ensure proper results. Please see the table below.
Kit part
| Sample volume
|
Standard
| Maximum
|
NPH | 20 µL | 30 µL |
PITC | 10 µL | dry after each 20 µL |
Lipid | 10 µL | dry after each 20 µL |
Information for PITC and Lipid parts
After adding a sample volume of 20 µL to a filter of the test plate, include an additional drying step of 30 min. In theory, there is no maximum sample volume. However, for sample volumes larger than 30 µL, matrix effects may become predominant.