Maximum unknown sample volumes for MxP Quant 1000 kit?

Maximum unknown sample volumes for MxP Quant 1000 kit?

If you wish to add more sample volume on a filter of the test plate, please note that extra drying steps are required to ensure proper results. Please see the table below.

Kit part

Sample volume

Standard

Maximum

NPH

20 µL

30 µL

PITC

10 µL

dry after each 20 µL

Lipid

10 µL

dry after each 20 µL


Info
Information for PITC and Lipid parts
After adding a sample volume of 20 µL to a filter of the test plate, include an additional drying step of 30 min. In theory, there is no maximum sample volume. However, for sample volumes larger than 30 µL, matrix effects may become predominant.