Pressure manifold with MxP Quant 1000 kit
If
a pressure manifold is to be used with the MxP Quant 1000 kit, a special
adapter is required for drying steps, please see the Product number: 20981.
Do not use the adapter for pressure manifold for the elution of extracts.
Related Articles
Why do extracts appear cloudy?
For AbsoluteIDQ® p180, p400 HR and MxP® Quant 500 kits: When adding water to the methanolic sample extracts to make the LC plate, cloudiness can occur due to residues of pyridine and phenyl isothiocyanate from the derivatization solution. If that ...
Why the extracts did not go through the filters or the volumes in the capture plate are low/inconsistent?
Inconsistent or lack of volume in the filter plate may occur for several reasons. Below are some suggestions: Water or another solvent was used for making the extraction solvent instead of Methanol. Perform the following steps: Transfer the extracts ...
Maximum unknown sample volumes for MxP Quant 1000 kit?
If you wish to add more sample volume on a filter of the test plate, please note that extra drying steps are required to ensure proper results. Please see the table below. Kit part Sample volume Standard Maximum NPH 20 µL 30 µL PITC 10 µL dry after ...
Can I use speed vac for drying the test plate?
Officially biocrates does not recommend using a speed vac instead of a nitrogen drying system. The use of a speed vac for kit preparation procedures was never tested by biocrates. However and for sure, a speed vac can be used. If a speed vac is used, ...
How long will my kits remain stable when stored?
Expiration date for Quant 500 and Quant 500 XL kits: Storage at -20 °C: 3 months after shipping date Storage at -80 °C: 12 months after shipping date Expiration date for p180, p400 HR, Quant HR Xpress and Bile Acids kits: Storage at -20 °C: 6 months ...