Why the extracts did not go through the filters or the volumes in the capture plate are low/inconsistent?

Why the extracts did not go through the filters or the volumes in the capture plate are low/inconsistent?

Inconsistent or lack of volume in the filter plate may occur for several reasons. Below are some suggestions:

  1. Water or another solvent was used for making the extraction solvent instead of Methanol. Perform the following steps:
  1. Transfer the extracts back to the filter plate.
  2. Dry the samples under nitrogen.
  3. Dissolve again using pure methanol (volume according to extraction step in user manual).
  1. The pressure manifold was not set up correctly.
  1. Check that the pressure manifold settings are correct according to the instructions.
see document “Instructions using pressure manifolds for 96-well plates with biocrates kits”
  1. Put the plate back in the manifold and try again.
  2. If the problem persists, increase the low flow pressure up to 30 psi.
  1. Droplets of liquid remaining on the underside of the filter plate. It is normal if a small amount of liquid is on the underside of the filter. Internal standards will account for small volume discrepancies. Take care that there is no cross contamination into other wells when removing the top filter plate. If large droplets remain on the underside of the filter plate:
  1. Put the plate back into the pressure manifold and run at a higher pressure (30 psi).
  2. While the filter and capture plate are still secured together by tape, tap the plate lightly on the table to shake the droplet loose.
  1. It is also possible that some amount of volume was lost during the shaking step or when transporting the plate. You can check the wells with lower volume in the capture plate to see if the remaining volume is still in the top filter plate. If the top filter plate is empty, then it’s likely that the volume was spilled at some point during the preparation. Take extra care during shaking and carrying the plate that nothing is spilled when the wells are uncovered!

    • Related Articles

    • How long will my kits remain stable when stored?

      Expiration date for Quant 500 and Quant 500 XL kits: Storage at -20 °C: 3 months after shipping date Storage at -80 °C: 12 months after shipping date Expiration date for p180, p400 HR, Quant HR Xpress and Bile Acids kits: Storage at -20 °C: 6 months ...
    • Why do extracts appear cloudy?

      For AbsoluteIDQ® p180, p400 HR and MxP® Quant 500 kits: When adding water to the methanolic sample extracts to make the LC plate, cloudiness can occur due to residues of pyridine and phenyl isothiocyanate from the derivatization solution. If that ...
    • Does the whole kit plate need to be prepared at once?

      A kit must be prepared all at once, as it is sealed under nitrogen to prevent degradation. Kits are for single use only.
    • What happens if I use the kit after it expires?

      All kit components requiring a refrigeration are subject to expiration (light blue box). These are the Vial Box and the Kit Plate. After passing the expiration date, the kit can still be used. However, the proper performance cannot be guaranteed. ...
    • Why are there signals in the blank or in the zero samples, why are analytes in the zero samples above LOD?

      The blank well A1 does not contain any analytes or internal standards. The zero samples only contain internal standards, but no analytes. If there are still peaks the following may have occurred: A sample or internal standard mix was accidentally ...