What could cause a loss in sensitivity of absence of peaks?

What could cause a loss in sensitivity of absence of peaks?

Please be aware that performance-related issues that cannot be resolved by instrument cleaning are not within biocrates’ responsibility and must be handled by the user or the LC-MS engineer. Based on our experience, however, we are happy to share some suggestions:

Sample preparation:
  1. Make sure the testmix/test sample has been reconstituted or diluted correctly.
  2. Make sure there was no air in the pipette tip when transferring sample to another vial.
  3. Make sure the testmix/test sample used was not expired.
Instrumental setup: double-check all method parameters and make sure the analytical column and mobile phases were prepared and installed correctly.

Connections: check for leaks and retighten all connections.

Injection needle: the injection needle might not be correctly adjusted or draws air. Adjust positioning or replace the needle.

Injector: check for leaks or contaminations in the injector valve and rotor seal. Clean or replace the valve.

Divert valve: set to waste instead of MS. Check setting at valve and in method.

ESI electrode: leak, contamination, or defective. Reassemble ESI probe, clean with different solvents at high flow rate, or replace ESI needle.

MS: perform instrument front-end cleaning (Orifice, QJet, quadrupole, ion transfer tube etc.). Tune and calibrate the instrument.


    • Related Articles

    • What should FIA peaks look like and how can the peak shape be improved?

      The best peak shapes are obtained by avoiding unnecessary dead volumes (caused by valves or connection parts) and connecting the injector directly to the mass spectrometer. For this, red PEEK tubing is recommended. The FIA peaks must be located ...
    • How to choose correct RT when there are multiple peaks?

      WebIDQ contains a tool Auto peak picking for defaults for automatically determine the retention time of each metabolite. It will set the same default retention time ranges for each sample based on the quality controls (QCs). After applying Auto peak ...
    • How should the LC-MS system be maintained for optimal performance?

      Important cleaning notes: Frequent cleaning actions and rinsing methods are strongly recommended to minimize the risk of instrument contaminations or performance loss. For optimal time-efficiency, clean and rinse the instruments routinely between the ...