If you wish to add more sample volume on a filter of the test plate, please note that extra drying steps are required to ensure proper results. Please see the table below. Kit part Sample volume Standard Maximum NPH 20 µL 30 µL PITC 10 µL dry after ...
The blank well A1 does not contain any analytes or internal standards. The zero samples only contain internal standards, but no analytes. If there are still peaks the following may have occurred: A sample or internal standard mix was accidentally ...
If a pressure manifold is to be used with the MxP Quant 1000 kit, a special adapter is required for drying steps, please see the Product number: 20981. Do not use the adapter for pressure manifold for the elution of extracts.
All kit components requiring a refrigeration are subject to expiration (light blue box). These are the Vial Box and the Kit Plate. After passing the expiration date, the kit can still be used. However, the proper performance cannot be guaranteed. ...