Why are the retention times shifted?

Why are the retention times shifted?

If the retention times are unstable across a plate or shifting over time:
- Make sure the mobile phases are purged and there are no bubbles in the system.
- Make sure no air is drawn or injected by the autosampler.
- Make sure the entire LC-MS system is well conditioned and the column is sufficiently equilibrated.

Otherwise, check the following parts with your instrument's vendor:

- Degasser
- Mixing cahmber (needs to be replaced)
- Intake/outlet valve at pump head
- Seals at pump head
- Replace the analytical column (it may have reached the end of its lifespan)
    • Related Articles

    • Unstable retention times

      Retention times (RTs) must be stable across multiple injections. Retention time shift Reproducible retention time Retention time fluctuations during a measurement are usually caused by the LC system. The most common causes are 1. Pump malfunctions, ...
    • How should the LC-MS system be maintained for optimal performance?

      Important cleaning notes: Frequent cleaning actions and rinsing methods are strongly recommended to minimize the risk of instrument contaminations or performance loss. For optimal time-efficiency, clean and rinse the instruments routinely between the ...
    • FIA blank contaminations

      The contaminations could be eluting from the autosampler. To reduce the contaminations in the FIA blank injections, perform the following procedure: Install strong wash (equal parts MeOH:ACN:Water:IsoPropanol) on the FIA line and flush for 20 min at ...