Why are there signals in the blank or in the zero samples, why are analytes in the zero samples above LOD?

Why are there signals in the blank or in the zero samples, why are analytes in the zero samples above LOD?

The blank well A1 does not contain any analytes or internal standards. The zero samples only contain internal standards, but no analytes. If there are still peaks the following may have occurred:

  1. A sample or internal standard mix was accidentally loaded.
  2. Spillover when loading the extraction solvent onto the kit plate, e.g. the dispensing speed of the repeater was too high.
  3. Spillover when shaking the plate after loading the extraction solvent.
  4. Cross-contamination by residues at the bottom of the filter plate when removing the filter plate from the capture plate.
  5. Cross-contamination when transferring the extracts to make the LC and FIA plates.
  6. To make sure the contamination comes from the plate and not from the system, inject a blank from an autosampler vial to cross-check.

    • Related Articles

    • What is the runtime per sample and kit?

      Run time per sample / min Run time per kit / h
    • Maximum unknown sample volumes for MxP Quant 1000 kit?

      If you wish to add more sample volume on a filter of the test plate, please note that extra drying steps are required to ensure proper results. Please see the table below. Kit part Sample volume Standard Maximum NPH 20 µL 30 µL PITC 10 µL dry after ...
    • What kind of samples can be measured with the kits?

      biocrates kits have been specifically validated for use with plasma and serum. Feasibility tests have been carried out with other sample matrices and protocols are available (see below). Basically, any samples containing endogenous metabolites can be ...
    • What kind of plasma anticoagulants are compatible with biocartes kits?

      biocrates kits were validated with EDTA and heparin plasma only. Citrate plasma may also be used, but performance may be reduced. The most important aspect is to ensure that all plasma samples within one study are collected in the same way and using ...
    • Why do extracts appear cloudy?

      For AbsoluteIDQ® p180, p400 HR and MxP® Quant 500 kits: When adding water to the methanolic sample extracts to make the LC plate, cloudiness can occur due to residues of pyridine and phenyl isothiocyanate from the derivatization solution. If that ...